Divesoft DNA 8460 Manuel de l'utilisateur - Page 10
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Divesoft DNA 8460. Divesoft DNA 8460 18 pages.
11/24/23, 1:33 PM
2. Select the calibration method AIR or OXYGEN.
3. Allow the calibration gas to flow into the analyzer and wait for
the sensor voltage to stabilize.
4. Press CONFIRM.
NOTE: Ensure that the sensor voltage value is within the sensor
manufacturer's parameters when exposed to the gas. DIVESOFT
oxygen sensors have an air voltage range of 7 - 13 mV. It is
advisable to verify that the sensor shows linear measurement
https://www.divesoft.com/en/manuals/8460-dna-user-manual?_ga=2.264787596.209083188.1700850446-662384236.1700850430
#8460 DNA User Manual
10/18