Arthur Grillo WR 200-F Beschreibung und Betrieb - Seite 3
Blättern Sie online oder laden Sie pdf Beschreibung und Betrieb für Messgeräte Arthur Grillo WR 200-F herunter. Arthur Grillo WR 200-F 16 Seiten. Heat meter
Application
The heat meter WR 200-F is used to detect the heat energy in closed heating or cooling
systems using a liquid heat carrier. The heat energy from the current heat output according
to the formula
P = Q · ∆T · k
calculated integrated over time and summed on a 9-digit counter.
The flow rate is also added up to another counter.
The heat coefficient k is calculated for water in the range of 0 ..
250 ° C as a function of flow and return temperature and the location of the flow measure-
ment. Basis for calculating are material properties for water according IAPWS-IF97.
For other heat carriers (e.g. Water-brine mixtures or thermal oil) can be considered equiva-
lent thermal coefficients in the unit.
P = heat output in kW
Q = Flow in m3/h
∆T = Temperature difference between flow-
and return temperature in °C
k
= Thermal coefficient in kWh / m
3
/h
3
·°C
3