BATTERY MODULE REMOVAL

Battery Module Removal
Follow these instructions for proper removal and storage of a the Battery
Module from the system.
1. Turn system Off with power
switch.
2. Disconnect the Power Cord
from the power supply.
3. Remove the Rear Panel and
disconnect the Power Cord from
the system.
4. Remove the Front Access Panel
and disconnect all cables and
lead wires.
5. Lift the system from the cart, if
equipped, and place upside
down on a level working surface
with a clean cloth or cushion
under the display lens to prevent
damage or scratches.
WARNING
Disconnect the system from the power source (wall outlet) before
attempting any maintenance, installation, removal or replacement
procedures to prevent electrical shock and possible damage to system.
REAR ACCESS PANEL
DISCONNECT
POWER CORD
6. Remove the four mounting
screws from the Module
securing it to the system using a
#1 Phillips Screwdriver.
7. Carefully lift the Module up and
disconnect the ribbon cable from
the system.
8. Store the module and the
associated accessories in an
area with temperatures
between 15° C and 40° C
(59° F and 104° F), and with
Relative Humidity ranging from
30%-60%.
9. Remove the extended Front
Access Panel, if no other Module
is to be installed, and reinstall
the low profile Front Access
Cover using a #1 Phillips
Screwdriver.
11
Intelect Advanced Battery Module