9. Special function Service Menu

Function of service Menu
Calibration
Display contrast
Error log
Device runtime
Entering the service menu:
Press the menu key until Service Menu
shows up (menu 3).
Press
for
is flashing on the display.
The display will change and show:
Entering code.
For the code press: 1 3 2 4
Now you see the symbols of the service menu
for 5 seconds,
9.1 Calibration
ATTENTION!
Before you do a calibration switch the device
OFF and ON.
Adjust a maximum femur length, a minimum lower leg
length, hip axis in smallest position and the middle
position of the foot rotation.
Press the symbol calibration
Display:
Press + or - to move the ARTROMOT®-K4 to 0
degrees.
Press START, the calibration starts automatically. The
device will reach both maximum points and move
between –10 bis 125 degrees with different speed.
Wait until the ARTROMOT®-K4 stops.
If the calibration was succesful the device stops at
0 degrees and show following symbols on the display:
Display:
Press STOP twice to leave the service menu.
Finally, a safety and function test has to be performed
(see chapter 11).
9.2 Display contrast
Press the symbol display contrast
Display:
Press + or - to set up the requested display contrast.
You can set the display contrast from 0 – 100%.
Press STOP twice to save the settings and leave the
service menu.