Before the chamber is shipped to a different location, all water must be emptied from the water reservoir.
The water can be emptied by tilting the chamber and emptying all the water into a bucket, through the
filling tube.
Make sure the unplug the chamber before filling or emptying water.
Setup
Before the H300 can be used, it needs to have sufficient water and desiccant for proper operation. The
chamber comes from the factory with the desiccant preloaded, but without any water.

Filling the Water Reservoir

Use only distilled water in the water chamber.
distilled (e.g. spring water). The impurities (salts) in regular water affect the vapor pressure and will limit
the maximum humidity the chamber can reach.
The H300 chamber should be filled with distilled water when no water is visible in the filling tube.
To fill the water reservoir follow the simple procedure below:
Unplug the H300 from the power source.
Unscrew the plug at the top of the filling tube (shown below).
H300 Temperature Humidity Chamber
Do not use tap water or bottled water that is not