MEASURING THE DENSITY OF A LIQUID

6.
Assemble the kit as described in the procedure below.

6.1. Assembling the Kit:

Assembled view
Thermometer clamp
1.
For the procedure up to the setting of the
beaker stand, refer to the assembly procedure
for each balance model described in "5.
MEASURING THE DENSITY OF A SOLID".
2.
Attach the thermometer clamp to the beaker
and
insert
thermometer clamp.
3.
Place the beaker on the beaker stand.
4.
Connect the float to the float hook using the
wire. The length of wire must be about 50 mm.
5.
Hook the float hook on the density pan stand.
6.
When the value displayed on the balance is
stable, press the RE-ZERO key to set the
displayed value to zero.
Now the balance is ready for density measurement.
Thermometer
Beaker
Beaker stand
the
thermometer
Float hook
Wire
Float
Density pan stand
Balance weight (depends on the balance model)
into
the
14
The volume of the
float is printed.
Example: 10.00 g/cm
Approx. 50 mm
3