9.
Install the new Sight Glass Assembly by inserting the
bottom tube through the copper coil and pushing
the two remaining tubes over the clamp arm toward
the back of the instrument.
10. Install the new tubes over their correlating
connections. Then, install new zip ties.
Notes: If pairing proves difficult, users may use a
blunt tool to slightly stretch the tubing very
carefully.
There are two types of tubing in this assembly.
Ensure not to crinkle the inner linings during
handling and installation.
ANKOM Technology • 2052 O'Neil Road •
Sight Glass Assembly (X85)
XT15/XT15I
https://www.ankom.com/contact/technical-services
• Phone: 315-986-8090
#021
Service Procedure
Revised: 11/1/22