6.
Remove all three tubes from their connections.
7.
Return to the front of the instrument and take hold
of the sight glass assembly by the silver plate.
There is a notch to accommodate your hand.
Pull the assembly out and away from the
instrument.
8.
Select the new Sight Glass Assembly (X85)
Note: Be sure not to bend or twist the tube, as kinks
can cause damage to the line.
ANKOM Technology • 2052 O'Neil Road •
Sight Glass Assembly (X85)
XT15/XT15I
https://www.ankom.com/contact/technical-services
Tubes
• Phone: 315-986-8090
#021
Service Procedure
Revised: 11/1/22
pg. 3 of 5