Ankom XT15I Процедура обслуживания - Страница 3

Просмотреть онлайн или скачать pdf Процедура обслуживания для Лабораторное оборудование Ankom XT15I. Ankom XT15I 5 страниц. Extractor sight glass assembly

6.
Remove all three tubes from their connections.
7.
Return to the front of the instrument and take hold
of the sight glass assembly by the silver plate.
There is a notch to accommodate your hand.
Pull the assembly out and away from the
instrument.
8.
Select the new Sight Glass Assembly (X85)
Note: Be sure not to bend or twist the tube, as kinks
can cause damage to the line.
ANKOM Technology • 2052 O'Neil Road •
Sight Glass Assembly (X85)
XT15/XT15I
https://www.ankom.com/contact/technical-services
Tubes
• Phone: 315-986-8090
#021
Service Procedure
Revised: 11/1/22
pg. 3 of 5