Electromagnetic emissions:
I TECH LITE is intended for use in the electromagnetic environment specified below. The I TECH LITE should only be used in such an
environment.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000‐3‐2
Voltage fluctuation/ flicker
emissions
IEC 61000‐3‐3
Electromagnetic Immunity:
I TECH LITE is intended for use in the electromagnetic environment specified below. The I TECH LITE should only be used in such an
environment.
Immunity test
6kV contact
Electrostatic discharge
8kV air
(ESD) EN61000‐4‐2
2kV for power supply
Burst/Fast transient
EN 61000‐4‐4
lines
1kV differential mode
Surge
2kV common mode
EN 61000‐4‐5
Immunity test
5%Ut
Voltage dips, short
interruptions and
(95% dip in U
voltage variation on
cycle
power supply input
lines.
40% Ut
(60% dip in U
70% U
(30% dip in U
cycle
5% U
EN 61100‐4‐11
(95% dip in U
Power frequency
(50/60 Hz)
magnetic field
3A/m
EN 61100‐4‐8
I TECH LITE is intended for use in the electromagnetic environment specified below. The I TECH LITE should only be used in such an
environment.
USER MANUAL Rev 25/05/2023
15. ACCOMPANYING DOCUMENTS
Group 1
I TECH LITE uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
Class A
I TECH LITE is suitable for use in all establishment including domestic establishments
Conforms
and those directly connected to the public low‐voltage power supply.
Conforms
Test level
EN 60601‐1‐2
6kV
8kV
2kV for power supply
lines
1kV differential mode
2kV common mode
Test level
ICE 60601
5%Ut
) for 0,5
(95% dip in U
t
cycle
40% Ut
) for 5 cycle
(60% dip in U
v
cycle
t
) per 25
70% U
t
(30% dip in U
cycle
t
5% U
) for 5 sec
t
3A/m
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]
I TECH LITE
Page 13 of 16
Compliance level
Floor should be wood, concrete, or ceramic tile. If
floor is covered with synthetic material, the relative
humidity should be at least 30%
Quality of the Main power source should be that of a
typical commercial or hospital environment
Quality of the Main power source should be that of a
typical commercial or hospital environment
Compliance level
Quality of the Main power source should be that of a
typical commercial or hospital environment If the
) for 0,5
t
user of the I TECH LITE requires continued operation
during power main interruptions, it is recommended
than the I TECH LITE be powered from an
uninterruptible power supply or battery.
) for 5
v
t
) per 25
t
t
(95% dip in U
) for 5
t
sec
Power frequency magnetic fields should be at a
typical commercial or hospital establishment level.
Electromagnetic environment ‐ guidance
Electromagnetic environment ‐ guidance