Electromagnetic emissions: 
I TECH LITE is intended for use in the electromagnetic environment specified below. The I TECH LITE should only be used in such an 
environment.  
RF emissions 
CISPR 11 
RF emissions 
CISPR 11 
Harmonic emissions 
IEC 61000‐3‐2 
 
Voltage fluctuation/ flicker 
emissions 
IEC 61000‐3‐3 
Electromagnetic Immunity: 
I TECH LITE is intended for use in the electromagnetic environment specified below. The I TECH LITE should only be used in such an 
environment.  
Immunity test 
6kV contact 
Electrostatic discharge  
8kV air 
(ESD) EN61000‐4‐2 
2kV for power supply 
Burst/Fast transient  
EN 61000‐4‐4 
lines  
 
1kV differential mode 
Surge  
2kV common mode 
EN 61000‐4‐5 
Immunity test 
5%Ut 
Voltage dips, short 
interruptions and 
(95% dip in U
voltage variation on 
cycle 
power supply input 
 
lines. 
40% Ut 
 
(60% dip in U
 
 
 
70% U
 
(30% dip in U
 
cycle 
 
 
5% U
 
EN 61100‐4‐11 
(95% dip in U
Power frequency 
 
(50/60 Hz) 
 
magnetic field 
3A/m 
 
EN 61100‐4‐8 
I TECH LITE is intended for use in the electromagnetic environment specified below. The I TECH LITE should only be used in such an 
environment. 
USER MANUAL Rev 25/05/2023 
 
15. ACCOMPANYING DOCUMENTS 
Group 1 
I TECH LITE uses RF energy only for its internal function. Therefore, its RF emissions 
are very low and are not likely to cause any interference in nearby electronic 
equipment. 
Class B 
 
 
Class A 
I TECH LITE is suitable for use in all establishment including domestic establishments 
Conforms 
and those directly connected to the public low‐voltage power supply. 
Conforms 
Test level 
EN 60601‐1‐2 
6kV  
8kV  
2kV for power supply 
lines  
 
1kV differential mode 
2kV common mode 
Test level 
ICE 60601 
5%Ut 
) for 0,5 
(95% dip in U
t
cycle 
 
40% Ut 
) for 5 cycle 
(60% dip in U
v
cycle 
 
 
t
) per 25 
70% U
t
(30% dip in U
cycle 
 
5% U
) for 5 sec 
t
3A/m 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
I TECH LITE 
Page 13 of 16 
Compliance level 
Floor should be wood, concrete, or ceramic tile. If 
floor is covered with synthetic material, the relative 
humidity should be at least 30% 
Quality of the Main power source should be that of a 
typical commercial or hospital environment 
Quality of the Main power source should be that of a 
typical commercial or hospital environment 
Compliance level 
Quality of the Main power source should be that of a 
typical commercial or hospital environment If the 
) for 0,5 
t
user of the I TECH LITE requires continued operation 
during power main interruptions, it is recommended 
than the I TECH LITE be powered from an 
uninterruptible power supply or battery. 
) for 5 
v
 
t
) per 25 
t
(95% dip in U
) for 5 
t
sec  
Power frequency magnetic fields should be at a 
typical commercial or hospital establishment level. 
 
 
Electromagnetic environment ‐ guidance 
Electromagnetic environment ‐ guidance