The micro motor cable has limited durability and is considered a consumable product.  The durability of the cable is largely dependent 
upon use and treatment, i.e., handling, reprocessing and frequency of use.  
Caution must be exercised when connecting or disconnecting the micro motor and foot control cable from the device to avoid damage 
to the cable. 
Periodically inspect the micromotor cord for damage prior to use. If there is damage noted, contact the Manufacturer. 
LUBRICATING: 
Do not oil or lubricate the micromotor. 
RUBBER O‐RING: 
If necessary, substitute. Contact the Manufacturer 
1.Do not attempt to disassemble the motor or motor connector. 
2.Do not oil or lubricate the motor. 
3. Do not attach a handpiece to the motor while the motor is running. 
4.Do not bend motor cord sharply. 
The Micromotor is sensitive to shock. Do not drop or impact micromotor against hard surface. 
Failure to comply with any of the above instructions may void your warranty. 
 
 
Manufacturer 
ATS Dental – Pistoia – Italy  
Model 
I TECH LITE 
Dimensions 
185x160x100 
Materials 
ABS 
Micromotor 
Brushless 40000/30000 
Noise 
< 65 dBA 
Supply Voltage 
24‐30 VDC 
Not suitable for use in presence of flammable anesthetics or oxygen. 
Conditions of use: Temp +18°C/+40°C (+64°F/+140°F) RH < 80% 
Shipping and Storage conditions: Temp +5°C/+65°C (+41°F/+149°F) RH < 20‐95% non‐condensing. 
Life of the device: When used as prescribed = 5 years. No warranty claim may be made if the failure happens sooner or later than 
indicated, because of the frequency of use, sterilization and maintenance. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
USER MANUAL Rev 25/05/2023 
 
12.TECHNICAL DATA 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
I TECH LITE 
Page 11 of 16 
11.Maintenance 
 
Nominal Power 
Frequency 
Protection 
Type 
Class MD 
Console 
Micromotor 
and Pedals 
 
 
100 VA 
50/60 Hz 
Class 1 
BF 
IIa rule 9 in Annex IX 93/42/EEC 
IP21 
IP21