Unplug the main unit from power source before cleaning or disinfecting. Do not use directly on liquids and sprays. Never immerse the 
unit or its parts in liquid.  
External  surface  of  the  console  may  be  cleaned  with  a  damp  cloth  or  disinfectant  wipe.  The  housing  is  not  waterproof.  Do  not  use 
ultrasonic or steam sterilizers. External surface of the micro motor may be cleaned with a damp cloth or disinfectant wipe. Do not place 
micro motor in ultrasonic sterilizers. Motor Holder and Irrigation Bag Hanger Rod may be cleaned with a damp cloth or disinfectant wipe.   
The exterior of the foot controls may be cleaned by wiping with a soft cloth moistened with mild detergent or disinfecting solution. 
 
WHAT CAN BE STERILIZED: 
"REF" ITL500 
Micromotor Assembly (Motor‐Cable‐Connector) 
Motor Holder 
Irrigation Bag Hanger Rod 
 
 
 
When treating patients who may have an acute, critical infectious disease, be sure to observe the hygienic measures cited in applicable 
publications and reports. If possible, use suitable disposable products to avoid the transmission of critical pathogens. These protect the 
user, the patient, and all participants in the surgery. A microbiologically appropriate coolant liquid must be used. Only use prescribed 
isotonic saline solution NaCl 0.9 (identified as an infusion solution; follow the instructions on the package insert) for cooling and 
spraying wounds. Not all autoclaves can reach 134° C. Not all autoclaves draw a pre‐vacuum. Please refer to your autoclave 
manufacturer for specific sterilization instructions. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
USER MANUAL Rev 25/05/2023 
 
10.  Cleaning and Sterilization 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
I TECH LITE 
Page 10 of 16 
STERILIZATION METHOD: 
Sterilization with steam or chemical vapor 
Sterilization time :18 minutes at 134°C/2 bar. 
Do not exceed 138° C or 275° F 
Cooling down time: 2 hours 
Working temperature: 40°C 
Max sterilization cycles: 500. 
Do not submerge in any solutions. 
Do not use ultrasonic cleaners. 
Do not sterilize with Ethylene Oxide