Unplug the main unit from power source before cleaning or disinfecting. Do not use directly on liquids and sprays. Never immerse the
unit or its parts in liquid.
External surface of the console may be cleaned with a damp cloth or disinfectant wipe. The housing is not waterproof. Do not use
ultrasonic or steam sterilizers. External surface of the micro motor may be cleaned with a damp cloth or disinfectant wipe. Do not place
micro motor in ultrasonic sterilizers. Motor Holder and Irrigation Bag Hanger Rod may be cleaned with a damp cloth or disinfectant wipe.
The exterior of the foot controls may be cleaned by wiping with a soft cloth moistened with mild detergent or disinfecting solution.
WHAT CAN BE STERILIZED:
"REF" ITL500
Micromotor Assembly (Motor‐Cable‐Connector)
Motor Holder
Irrigation Bag Hanger Rod
When treating patients who may have an acute, critical infectious disease, be sure to observe the hygienic measures cited in applicable
publications and reports. If possible, use suitable disposable products to avoid the transmission of critical pathogens. These protect the
user, the patient, and all participants in the surgery. A microbiologically appropriate coolant liquid must be used. Only use prescribed
isotonic saline solution NaCl 0.9 (identified as an infusion solution; follow the instructions on the package insert) for cooling and
spraying wounds. Not all autoclaves can reach 134° C. Not all autoclaves draw a pre‐vacuum. Please refer to your autoclave
manufacturer for specific sterilization instructions.
USER MANUAL Rev 25/05/2023
10. Cleaning and Sterilization
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]
I TECH LITE
Page 10 of 16
STERILIZATION METHOD:
Sterilization with steam or chemical vapor
Sterilization time :18 minutes at 134°C/2 bar.
Do not exceed 138° C or 275° F
Cooling down time: 2 hours
Working temperature: 40°C
Max sterilization cycles: 500.
Do not submerge in any solutions.
Do not use ultrasonic cleaners.
Do not sterilize with Ethylene Oxide