Freedom Innovations Kinnex 2.0 Manual de instrucciones - Página 6

Navegue en línea o descargue pdf Manual de instrucciones para Equipamiento médico Freedom Innovations Kinnex 2.0. Freedom Innovations Kinnex 2.0 11 páginas. Prosthetist
También para Freedom Innovations Kinnex 2.0: Manual del paciente (2 páginas), Manual de instalación (2 páginas), Manual de referencia (2 páginas)

Freedom Innovations Kinnex 2.0 Manual de instrucciones

LED Button Quick Reference Sheet

Each button on Kinnex incorporates one or more indicator
lights to indicate the status of Kinnex. The table below
lists what the indicator lights mean.
Indicator Light
Solid green
Flashing green
Solid red
Flashing red
Solid white
Flashing white
Solid blue
Flashing blue
Alternating red and
white
Alternating red and
blue.
10
Location
Battery Button
Battery Button
Battery Button
Battery Button
Change Footwear Button
Change Footwear Button
Bluetooth Button
Bluetooth Button
Battery and
Change Footwear
Buttons
Battery and
Bluetooth Buttons
Freedom Innovations |
Meaning
Battery is fully charged.
Charging (plugged in):
Battery is charging.
Plugged in (discharging):
Battery charge level has
dropped below 50%.
Battery charge level
is low.
Battery is almost fully
depleted. Once fully depleted,
Kinnex will go into a low
stiffness mode.
Manual Lock is engaged.
Manually adjusting
footwear.
Connected in Clinician Mode.
Ready for Clinician Mode
connection.
User: Attempt to
change footwear failed.
Prosthetist: Sensor
calibration failed.
Kinnex system error.
Kinnex 2.0
| Instructions for Use
Action
Kinnex is fully charged and
ready to use.
Charging (plugged in):
Continue charging battery
until solid green light
for full charge.
Unplugged (discharging):
Charge the battery in the
next few hours.
Charge battery as soon
as possible.
User will notice
reduced resistance.
Take caution performing
normal activities.
Kinnex is locked in
desired position.
Stand evenly weighted
and wait 5 seconds.
Clinician can configure
Kinnex. To exit, press
Bluetooth Button for 3
seconds until the blue
indicator light turns off.
Clinician can connect to
Kinnex. To exit, press
Bluetooth Button for 3
seconds until the blue
indicator light turns off.
User: Repeat change
footwear process.
Prosthetist: Repeat
sensor calibration process.
Contact Prosthetist.
11