Apogee SP-214 Manual del usuario - Página 7
Navegue en línea o descargue pdf Manual del usuario para Accesorios Apogee SP-214. Apogee SP-214 18 páginas. Pyranometer
También para Apogee SP-214: Manual del usuario (19 páginas)
Spectral Response
Temperature Response
Spectral response estimate of Apogee silicon-cell
pyranometers. Spectral response was estimated
by multiplying the spectral response of the
photodiode, diffuser, and adhesive. Spectral
response measurements of diffuser and adhesive
were made with a spectrometer, and spectral
response data for the photodiode were obtained
from the manufacturer.
Mean temperature response of four Apogee
silicon-cell pyranometers. Temperature response
measurements were made at approximately 10 C
intervals across a temperature range of
approximately -10 to 50 C under sunlight. Each
pyranometer had an internal thermistor to
measure temperature. At each temperature set
point, a reference blackbody pyranometer was
used to measure solar intensity.
7