Apogee SP-214 Instrukcja obsługi - Strona 7
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Akcesoria Apogee SP-214. Apogee SP-214 18 stron. Pyranometer
Również dla Apogee SP-214: Instrukcja obsługi (19 strony)
Spectral Response
Temperature Response
Spectral response estimate of Apogee silicon-cell
pyranometers. Spectral response was estimated
by multiplying the spectral response of the
photodiode, diffuser, and adhesive. Spectral
response measurements of diffuser and adhesive
were made with a spectrometer, and spectral
response data for the photodiode were obtained
from the manufacturer.
Mean temperature response of four Apogee
silicon-cell pyranometers. Temperature response
measurements were made at approximately 10 C
intervals across a temperature range of
approximately -10 to 50 C under sunlight. Each
pyranometer had an internal thermistor to
measure temperature. At each temperature set
point, a reference blackbody pyranometer was
used to measure solar intensity.
7