DayTronic 3000PLUS Manual de instrucciones - Página 6

Navegue en línea o descargue pdf Manual de instrucciones para Instrumentos de medida DayTronic 3000PLUS. DayTronic 3000PLUS 20 páginas. Panel meter, frequency conditioner module
También para DayTronic 3000PLUS: Manual de instrucciones (20 páginas)

DayTronic 3000PLUS Manual de instrucciones
cially noisy. Responding only to the modulated carrier
frequency, the module rejects extraneous voltages that
can cause errors in DC systems, particularly when
there is a need to "blow up" a portion of the transducer
range.
The conditioner's advanced analog design directly
addresses the problem of measurement inaccuracy in
industrial environments of high electromechanical
noise. Exceptional signal stability and accuracy over a
remarkably wide range of sensor inputs are achieved
through
• regulated, remotely sensed AC excitation
• precise linearity / symmetry correction and signal
phase adjustment via software commands
• high-stability amplification
• configurable low-pass active filtering
• "absolute" and "two-point" software-based
calibration
• effective signal isolation & ESD protection
Plugging into the rear of the 3000PLUS meter, the 5D78
connects directly to its source strain gage sensor via
simple screw-terminal pinout (see Fig. 3). High output
accuracy over a wide temperature range guarantees
repeatable sensor signal integrity. For steady indication
and smooth, dependable control action, the conditioner
provides a true average value of the measured variable,
even in the face of substantial dynamic content.
The 5D78 features
• Powerful user-selectable low-pass active filter-
ing for removal of unwanted high-frequency mea-
surement-signal components and the elimination of
aliasing errors
• High noise rejection, eliminating errors from com-
mon-mode pickup and ground-loop coupling, with
1500 VAC isolation between input and output termi-
nals and between I/O and power supply / communi-
cations terminals
* Employing a low-capacitance cable will minimize the effects on
overall offset and span. See note (**) on page 1.4.
3000PLUS
5D78 I
WITH
1. I
NTRODUCTION
• High ESD immunity and extensive EMI protec-
tion further assure data integrity in harsh industrial
environments
Because of normal cable loading effects, it is a practical
necessity to calibrate any AC TORQUE SENSOR/
CABLE/INSTRUMENT system after installation, using a
known input standard. The "ABSOLUTE" CALIBRATION
method described in Sections 3.B and 4.E will normally
yield a very good first approximation for a sensor/
cable/5D78 system (depending, of course, on actual
cable length and capacitance).* However, absolute cal-
ibration of a 3000PLUS with installed Model 5D78
should always be followed by a conventional "in place"
TWO-POINT (DEADWEIGHT) CALIBRATION, for opti-
mum data integrity.
To perform initial ABSOLUTE CALIBRATION, you need
only use the front-panel menu or configuration software
to specify the desired relationship between the mod-
ule's measured engineering units and its ±5-VDC out-
put, given the full-scale input range for which it is
currently set and the source transducer's known full-
scale rating and sensitivity. A zero offset term may also
be entered, expressed either in engineering units or
millivolts.
TWO-POINT (DEADWEIGHT) CALIBRATION requires
the application of two independently and accurately
known values of input loading ("ZERO" and "SPAN"), fol-
lowing the entry of a phase shift adjustment. Activation
of a user-supplied shunt calibration resistor is provided
(for either a positive or negative upscale reading) for
use in the two-point method, if desired. Internal sym-
metry trimming is available for negative-domain slope
adjustments of up to ±2% of full scale, along with both
positive and negative midscale linearity correction.
M
NSTRUCTION
ANUAL V
. SB.3
3KP78 - 1.2