dynarex 12001 Manual del usuario - Página 3
Navegue en línea o descargue pdf Manual del usuario para Equipamiento médico dynarex 12001. dynarex 12001 8 páginas. Long-term care bed
TECHNICAL SPECIFICATIONS
Bed Height
Back Rest Angle
Knee Break Angle
Additional Manual Leg Lift
Trendelenburg/Reverse Trendelenburg
Auto Contour
Weight Capacity
Safe Working Load (SWL)
Sleeping Surface Dimension (Width × Length)
Caster Size
Gross Weight
Net Weight
Power Supply (V/PH/Hz)
Power Consumption (Max)
Number of Motors
Protection Class
Insulation Class
Duty Cycle
Battery backup
Operating Temperature
Storage Temperature
Operation & Storage Humidity
Operation & Storage Atmospheric Pressure
Operation & Storage Altitude
Noise Level
4
10.4"-24.4"
0°-70°
0°-37°
N/A
N/A
Yes
470 lb.
450 lb.
36" × 80"
4"
200 lb.
165 lb.
100-240V/1ph/50-60Hz
230 W
3
IPX4
II Type B
Interruption 10%, Max. 2 min./18min.
Yes, optional
40°F-105°F (5°C-40°C)
15°F-120°F (-10°C-50°C)
20%-80%
700-1060 hPa
≤ 3000 m
< 50db(a)
Long-Term Care Bed • D100 (Reorder No. 12001)
ASSEMBLY INSTRUCTIONS
1.
UNPACKING THE BED
1. Unpack the components from the shipping
cartons.
2. Remove any black zip ties holding the components
to the Bed Assembly. White zip ties should not be
removed.
Bed Assembly × 1
Mattress Retainer × 6
Allen Wrench (size 5 mm)
NOT INCLUDED
Long-Term Care Bed • D100 (Reorder No. 12001)
3. Upon unpacking your bed, the package should
contain all the components shown below.
4. Inspect the bed components for damage. If the
bed is damaged, do not use the bed and contact
the manufacturer.
Head/footboard
Wall Bumper Bar × 1
Insertion Plate × 4
M10 Hex Bolt × 8
Wrench × 1
5