iHealth BG5 Manual del usuario - Página 12

Navegue en línea o descargue pdf Manual del usuario para Medidor de glucosa en sangre iHealth BG5. iHealth BG5 16 páginas. Wireless smart gluco-monitoring system
También para iHealth BG5: Manual de inicio rápido (13 páginas), Manual de inicio rápido (2 páginas), Manual de inicio rápido (7 páginas)

iHealth BG5 Manual del usuario
ethoxyethyl dimethyl benzyl ammonium chloride as the
active ingredient, have been shown to be safe for use with
the BG5 Blood Glucose meter and lancing device.
You can purchase this product according to the following
information: visit the web site www.metrex.com or contact
Metrex at 800-841-1428 for product or technical information.
The meter and the lancing device are for single-patient use,
if user test 5-6 times every day, the devices may be cleaned
and desinfected 5-6 times per day, as a result, there will be
11000 cycles cleanings plus disinfections (1 cycle = 1
cleaning + 1 disinfection) for the meter and the device during
its 5 years life cycle.
1. After the test, the user should clean and wash hands, and
then put on the disposable glove.
2. Use the Cavi WipesTM (Metrex
EPA Reg. NO. 46781-8, EPA Est. No. 56952-WI-001) to
clean and disinfect the meter's housing and lens,.Take one
wipe to clean the meter, including its front and back for 10
seconds for cleaning until there is no soil on the surface.
Then disinfect the meter with another wipe and keep the
surface wet for 2 min
With the same method use the Cavi WipesTM (Metrex
RESERCH CORPORATION,EPA Reg. NO. 46781-8, EPA
Est. No. 56952-WI-001) to clean and disinfect the lancing
device.
®
RESERCH CORPORATION,
®
20
4. Take off the glove and the protective cover.
Note:
1. Each disinfection step requires a pre-cleaning step.
2. The users should wash hands thoroughly with soap and
water after handling the meter, lancing device, or test
strips.
3. The meter can only be cleaned and disinfected with the
disinfecting towelette on the surface of the device, users
can not insert the disinfecting towelette into the test strip
port and the metal contector, and else the performance
of the meter may be influenced.
4. Please refer to the following linkage for detail information:
"FDA Public Health Notification: Use of Fingerstick
Devices on More than One Person Poses Risk for
Transmitting Bloodborne Pathogens: Initial
Communication" (2010)
http://www.fda.gov/MedicalDevices/Safety/AlertsandNoti
ces/ucm224025.htm
"CDC Clinical Reminder: Use of Fingerstick Devices on
More than One Person Poses Risk for Transmitting
Bloodborne Pathogens" (2010)
http://www.cdc.gov/injectionsafety/Fingerstick-DevicesB
GM.html
5. If you have any questions you can call the number
+1-855-816-7705
6. Table: A list of the signs of potential deterioration
Physical and in performance
What do you see?
What to do?
The device does not work,for
Stop using and contact local customer services or
example the iOS device can't
place of purchase for assistance, or toll free service
begin testing when meter is
hotline: +1-855-816-7705
connect with iOS device or
when test strip is inserted into
the meter.
21