Quick Start Guide
3
Troubleshooting
3.1
Transmitter will not zero despite having no pressure
applied
CAUTION
Before continuing, ensure no pressure is applied to single-use
process connector.
Solution: Factory reset required
Recommended actions
6
1. Disconnect the process connector from the transmitter.
2. Attempt a zero (see
•
If the zero was successful with no process connector
attached, then reattach the transmitter with the clamp.
•
If the zero was unsuccessful, then perform a factory
reset by holding a magnet on the magnet point for
four seconds.
3. Perform the zero function again with the clamp applied.
•
If the zero procedure fails, then perform a factory reset
by holding a magnet on the magnet point for four
seconds.
•
If the factory reset is successful, then the LED will
illuminate for four seconds before returning to factory
calibration.
•
If this procedure fails, then contact your local Emerson
representative.