Troubleshooting

3.1 
Transmitter will not zero despite having no pressure
applied
 CAUTION
Before continuing, ensure no pressure is applied to single-use
process connector.
Solution: Factory reset required
Recommended actions
6
1. Disconnect the process connector from the transmitter.
2. Attempt a zero (see
If the zero was successful with no process connector
attached, then reattach the transmitter with the clamp.
If the zero was unsuccessful, then perform a factory
reset by holding a magnet on the magnet point for
four seconds.
3. Perform the zero function again with the clamp applied.
If the zero procedure fails, then perform a factory reset
by holding a magnet on the magnet point for four
seconds.
If the factory reset is successful, then the LED will
illuminate for four seconds before returning to factory
calibration.
If this procedure fails, then contact your local Emerson
representative.