RIFD TAG SETUP

Detecting a Tag
To carry out an accurate measurement, it is
important that you provide the UT5000 with
the correct velocity of sound in the medium
to be measured. The UT5000 contains a table
of values for the most common materials.
For more unusual materials, you may need to
enter the velocity manually.
Tag Alias
Use the on-screen keyboard to define the
name of the RFID Tag. Use the arrow keys to
highlight the required character, then press
the SEL key. Repeat this procedure until you
have entered the full name. Use the arrow
keys to highlight ACCEPT, then press the SEL
key.
Material Velocity
Select a predefined material and velocity from
the UT5000's internal table or define your
own. The value will be assigned to the RFID
Tag. This procedure has been described on
page 14.
RUGGED
AUTHORITY
Setup Alarms
Set the low and high alarms for thickness
measurements within the vicinity of this RFID
Tag.
To set an alarm, highlight the Low or High
alarm option, then press the SEL key. Use the
arrow keys to highlight the required digit, then
press the SEL key. Repeat this procedure until
you have entered the full value. Use the arrow
keys to highlight ACCEPT, then press the SEL
key.
The ALARM key on the main display provides a
shortcut to this menu option.
You can also set up RFID tag information
through the UT5000 mini app (see page 11).
TRANSDUCER
SETUP
Pulse Frequency
The pulse frequency determines the rate of
repeated measurements. To change the pulse
frequency:
Press the FREQ key, or
Press the SETUP key. Use the arrow keys to
highlight Pulse Frequency, then press the
SEL key
Choose from the following options:
Single shot
The UT5000 takes one measurement
1, 2, 3, 4, 5, 6, 7 or 8 Hz
Measurements are taken continuously at the
selected rate and continue until the STOP
key is pressed
Transducer Select
Identify the type of transducer probe, e.g.
4MHz transducer.
Multi Echo
The Multi Echo option allows averaging of
individual measurements for greater accuracy
on uneven surfaces.
UT5000 – Intrinsically Safe Ultrasonic Thickness Gauge
Choose from the following options:
Off
2,3 or 4
Use the arrow keys to highlight the required
option then press the SEL key.
Probe Zero Built In
Select this option to calibrate/zero the probe
using the UT5000's built-in calibration disc
(see page 9).
Probe Zero/Cal Double
Select this option to calibrate/zero the probe
using the UT5000's built-in calibration disc
(see page 9).
Measure Material Velocity
Select this option to estimate the material
velocity for a sample of known thickness.
17