Gamry Interface 5000 Manual de inicio rápido - Página 2

Navegue en línea o descargue pdf Manual de inicio rápido para Equipos de prueba Gamry Interface 5000. Gamry Interface 5000 4 páginas. Installing standard cable on reference 600+
También para Gamry Interface 5000: Instrucciones (2 páginas)

Gamry Interface 5000 Manual de inicio rápido
When possible, connect the Chassis Ground on the back of your
potentiostat to a known, good earth ground.
Connect the cell cable to the 200 Ω Calibration Cell included with your
instrument.
Place the dummy cell inside the Calibration
Shield, close the lid, and connect the black
floating-ground lead of your cell cable to the
Shield's grounding post.
Open Gamry Framework™. Select Experiment > Utilities > Calibrate Instrument
Choose your potentiostat. Select the radio button for calibration type Both, and click the OK button.
For newer Interface™ potentiostats, the AC
radio button is inactive.
After you confirm several instructional messages, the calibration proceeds automatically, and you are
notified if the calibration is successful.