Gamry Interface 5000 Manuel de démarrage rapide - Page 2
Parcourez en ligne ou téléchargez le pdf Manuel de démarrage rapide pour {nom_de_la_catégorie} Gamry Interface 5000. Gamry Interface 5000 4 pages. Installing standard cable on reference 600+
Également pour Gamry Interface 5000 : Instructions (2 pages)
When possible, connect the Chassis Ground on the back of your
potentiostat to a known, good earth ground.
Connect the cell cable to the 200 Ω Calibration Cell included with your
instrument.
Place the dummy cell inside the Calibration
Shield, close the lid, and connect the black
floating-ground lead of your cell cable to the
Shield's grounding post.
Open Gamry Framework™. Select Experiment > Utilities > Calibrate Instrument
Choose your potentiostat. Select the radio button for calibration type Both, and click the OK button.
For newer Interface™ potentiostats, the AC
radio button is inactive.
After you confirm several instructional messages, the calibration proceeds automatically, and you are
notified if the calibration is successful.