Avante Volumax VOS Manual de consulta rápida - Página 4
Navegue en línea o descargue pdf Manual de consulta rápida para Producto para el cuidado de mascotas Avante Volumax VOS. Avante Volumax VOS 4 páginas.
Avante Volumax VOS
Quick Reference Guide
ALARM ICON
CARTRIDGE FAULT
CARTRIDGE TYPE
(CONTINUOUS AND FLOW
RATE NUMERIC DISPLAY
TEMPERATURE DISPLAY
SHOWS DASHES (
TEMPERATURE NUMERIC
CANNULA FLOW RATES
High Flow
Low Flow
2601 Stanley Gault Pkwy, Suite 101
800-923-8413
502-244-6345
•
INDICATES
Cartridge and/or
DPC not
detected. Unit will
not run.
Gas bubbles in
water circulation.
Unit continues to
operate.
Cartridge
and/or DPC not
detected.
LO
HI
Gas supply pressure
outside 4-85 psi (28-
586 kPa) range. Unit
will not operate.
GAS SUPPLY
(FLASHING)
not be provided
GAS SUPPLY
from current gas
supply.
FLASHES)
Temperature
out of range.
GENERAL FAULT
- -
)
FLASHING
Temperature 2° >
set point
Temperature 2° <
DISPLAY FLASHES
set point
CARTRIDGE
Adult, Pediatric/Adult Small, Pediatric Small*
Premature, Neonatal, Infant, Intermediate
Infant, Solo, Pediatric Small*
Louisville, KY 40223 USA
•
FAX: 502-244-0369
•
CAUSE
RUN mode: faulty
sensor or cartridge
not detected.
Excessive gas
STANDBY mode:
missing cartridge.
Gas supply is discon-
nected or exhausted.
Inlet gas pressure too
rate.
Overheating or
temperature sensor
malfunction.
User enters set point
much lower than pre-
vious temperature.
Very low water tem-
perature after bag
replacement.
CANNULA TYPE
www.avantevet.com
ACTION
Disconnect patient. Remove disposable
patient circuit. Check cartridge
installation. Check sensor windows
are clean.
Disconnect
patient circuit including
water path, cartridge & delivery tube.
Remove disposable patient circuit.
Check cartridge installation.
Check gas supply and correct as
necessary.
Increase gas pressure or decrease
Cannot be corrected by user: disconnect
Silence alarm and wait for
temperature to drop.
Silence alarm and wait for
temperature to rise.
OPERATIONAL FLOW RATES
5-40 liters per minute (lpm)
1-8 liters per minute (lpm)