Quick Reference Guide
DISPOSABLE PATIENT CIRCUIT
•
Attach High or Low Flow Vapor Transfer Cartridge to Disposable Water Path
•
Attach patient delivery tube to Disposable Water Path
•
Open door and install Disposable Water Path into docking station
•
Hang sterile water bag
•
Wipe water spike with alcohol pad and insert into sterile water bag
•
•
Volumax VOS is ready for start up
START UP AND ADJUSTING PARAMETERS
•
Install oxygen sensor
•
•
•
Rotate the blue Setting Control Knob to illuminate display
•
Press in Setting Control Knob to select the parameter and rotate to adjust
the value
•
Press and release (do not hold) the Run/Standby button once to start.
A GREEN light indicates RUN mode (AMBER light indicates STANDBY mode,
No Flow)
•
CONNECT TO PATIENT
•
•
Place the cannula on the delivery tube
•
After cannula is attached, it is ready to be placed on patient
•
To activate the internal backup battery mode, the Run/Standby LED must
be green before disconnecting the power cord.
•
percentage for at least 15 minutes.
•
SHUT DOWN MODE
•
Press the Run/Standby button. Unit will enter Standby mode.
•
Clamp the water inlet tube and disconnect it by pulling out the spike.
•
Open the door, remove the Disposable Patient Circuit (includes delivery
tube & cartridge) by sliding it upwards out of the docking station
•
Discard all disposables according to hospital guidelines.
•
Disconnect unit from AC power.