avantor VWR VC100 Manual de instrucciones - Página 10
Navegue en línea o descargue pdf Manual de instrucciones para Controlador avantor VWR VC100. avantor VWR VC100 19 páginas. Controller vacuum
3.2.3. Instrument configuration
Gently touch the "SETTINGS" button to pop up the password soft
keyboard, enter the password (the default initial password is 6666), then
press "enter" key, confirm the password is correct and then pop up the
setting interface, as shown in the figure below.
Name
Pump Delay(ms)
Valve Delay(ms)
K B
,
P0(mbar)
Reset
After setting, click "OK" key to return to the startup interface.
Note:
Before clicking the " Reset" button, record the pressure sensor calibration
parameters (K and B). After restoring the default setting, write (K and
B) into the instrument, otherwise the pressure measurement accuracy will
be adversely affected.
Fig. 3.2.2
Function introduction
The delay time after opening or closing the vacuum
pump.
The delay time after opening or closing the solenoid
valve (inside the instrument).
Pressure sensor calibration parameters
Local atmospheric pressure value
Restore the factory settings
Table 5
10