System Overview
Figure 1-7 Control panel
Source Exhaust System
The source exhaust system is a safety feature that isolates the sample vapors and
exhaust products from the laboratory environment. The source requires that the
source exhaust system is properly connected and functioning. The exhaust pump
draws the solvent vapors from the enclosed source housing and delivers them to
a trap at the rear of the instrument chassis. It is recommended that these vapors
be passed through this trap and then vented to a fume hood or outside port.