AB Sciex API 5000 LC/MS/MS Hardware-Handbuch - Seite 10

Blättern Sie online oder laden Sie pdf Hardware-Handbuch für Laborausrüstung AB Sciex API 5000 LC/MS/MS herunter. AB Sciex API 5000 LC/MS/MS 16 Seiten.

AB Sciex API 5000 LC/MS/MS Hardware-Handbuch
System Overview
Figure 1-7 Control panel

Source Exhaust System

The source exhaust system is a safety feature that isolates the sample vapors and
exhaust products from the laboratory environment. The source requires that the
source exhaust system is properly connected and functioning. The exhaust pump
draws the solvent vapors from the enclosed source housing and delivers them to
a trap at the rear of the instrument chassis. It is recommended that these vapors
be passed through this trap and then vented to a fume hood or outside port.
10
VACUUM
POWER
STATUS
ON
I
FAULT
O
OFF
Spare Output B
F4
Heated Nebulizer (APCI)
F3
F2
Card Cage Blower
F1
Spare Output A
WARNING
(see Note
below)
RESET
.
Document Number: D1000092206 B