Running a Blank
Blanks must be run every 12 hours, after a Soak Cleaning has been performed, after a
reagent pack change, and after the instrument has been restarted. It is not necessary
to run blanks on days that samples are not run. Additionally, blanks need to be run after
self tests and Deep Cleaning.
1
Press OK to measure a Blank first
2
Run
Blank results are expired!
1. From the Home Screen, touch Measure
then Run. If no blank has been run in the
last 12 hours, an expired blank message
will appear. Touch OK to start the blank.
2. If no warning message appears, Accept
the blank and proceed to running patient
samples.
3. If blank values are high, a pop-up warning
will appear. Touch OK, to view the blank
results and the corresponding Warning
flag (top right of screen).
There are parameters with high
blank values! Check the User's Manual
Troubleshooting section for guidance
and then repeat the blank
If Warning flags are present, refer
to the troubleshooting section of this
guide or the Operator's Manual to resolve
the warning.