Warning: Do not attempt to charge any other battery pack.
Do not use the battery packs supplied with this unit in any other
device. Use of the Biomet
Stimulator System battery packs in any other device may cause
damage or malfunction to the battery pack and/or devices. Do not
short circuit, overcharge, crush, mutilate, nail penetrate, heat, reverse
the + or – terminals or disassemble the battery pack. Do not allow
metal objects to come into contact with the battery pack terminals.
These and any other abuses of the battery pack may cause serious
injury and/or burns. To ensure proper charging and safety, use only
the charger supplied with your device. Keep battery pack dry. This
battery pack must be disposed of properly. Disposal information
can be obtained by contacting the Rechargeable Battery Recycling
Corporation (RBRC) at 1-800-822-8837 in the US.

Step 4: Changing Battery Packs

Once daily, preferably at the same time every day to ensure treatment
is continued without interruption, patients should do the following:
A. D epress the battery door latch (1) and slide the battery door on
the back of the stimulator (2) and remove the depleted battery
pack (Figure 7).
B. Following the arrows, place
the depleted battery pack
into the battery charger cradle
for charging (Figure 8). A solid
orange light on the charger
cradle will illuminate indicating
a proper connection. If no
light appears on the charging
cradle an error is indicated.
If this occurs, try removing
the battery pack from the
charger cradle and reinserting
it. If the orange light does not
appear, contact Biomet.
C. O nce the charger cradle's
orange light turns off and a
solid green light appears, the
battery pack is fully charged.
Remove the battery pack
from the battery charger
cradle with a gentle lift (1, 2)
on the battery tab. (Figure 9)
and place the fully charged
battery pack into the Biomet
SpinalPak
Non-invasive Spine
®
Fusion Stimulator System in
order to commence treatment.
D. T here should always be one battery pack in the charger and one
battery pack installed in the stimulator at all times, ensuring a fully
charged battery pack every 24 hours as recommended.
SpinalPak
Non-invasive Spine Fusion
®
®
®
5
Figure 7
Figure 8
Figure 9