Dickson TH8P0 Manual - Página 2
Navegue en línea o descargue pdf Manual para Instrumentos de medida Dickson TH8P0. Dickson TH8P0 8 páginas. Temperature & humidity chart recorder, 8” (203mm)
También para Dickson TH8P0: Manual del usuario (9 páginas)
Option 1:
Set the chart time manually by inserting a coin into the groove in the chart hub and turning clockwise until the correct
hour (and day if applicable) on the chart is referenced to the timing arrow (just to the right of the pen tip on the dial).
Press Pen Home to move the pens back onto the chart.
Option 2:
The following feature should be used for fine adjustments only.
To adjust the chart time, press and hold Adjust-Up and Adjust-Down buttons located on the back of the unit next to
the dip switch (See Button Operations below).
The green LED will blink rapidly for about five seconds, and then the LED will remain solid green. While in this state
the Adjust-Up button will move the chart backward (counter clockwise) and the Adjust-Down button will move the
chart forward (clockwise).
Rotate the chart until the correct hour (and day if applicable) on the chart is referenced to the timing arrow. Once you
have set your chart, press the Pen Home button to exit Chart Adjust Mode. The unit will take one minute to exit the
Chart Adjust Mode once Pen Home is pressed. Press Pen Home to move the pens back onto the chart.
Place the TH8 recorder on a flat vibration-free surface. Be sure it is in a vertical position and level. For best performance and
longevity, the location should be a clean environment, free from dust and corrosive fumes. Do not exceed temperature
specifications. Wall Mount: Keyhole slots are provided on the TH8recorder for wall mounting.
Display Symbols
Channel indicator
Current Reading or
Flashing "ALRN"
when in alarm
Low Battery Indicator
On/Off
The On/Off key turns the unit on and off.
Pen Home (Down Arrow)
If the pens are located on the outside edge of the chart, press the Pen Home to move pen(s) to recording position. If the pens
are located on the chart, press the Pen Home key to move the pen(s) to the outside edge of the chart.
Alarm Optional
Pen 1 = Red Pen
Pen 2 = Blue Pen
1. To set the alarm, make sure unit is on and press and hold the Alarm button. The LED will blink red rapidly for about 5 s,
then the LED will turn solid green. Release the Alarm button and the LED will turn solid red. At this point the display will
show "On" or "Off".
Pressing either the Adjust-Up or Adjust-Down buttons, located on the back of the unit next to the dip switch, will toggle
the alarm on or off.
Display when
setting alarms
Unit of measure for
channel displayed,
°F or °C