Alpha Moisture Systems DS1200-AMT Manual de instalación, puesta en marcha y funcionamiento - Página 6

Navegue en línea o descargue pdf Manual de instalación, puesta en marcha y funcionamiento para Instrumentos de medida Alpha Moisture Systems DS1200-AMT. Alpha Moisture Systems DS1200-AMT 18 páginas. Dewpoint analysis system

Alpha Moisture Systems DS1200-AMT Manual de instalación, puesta en marcha y funcionamiento
Note : - Do not install the sensor at this time. Wait until the commissioning stage as described later
4.3

Alarm Cable

Make the appropriate connections, noting the normally open and normally closed relay contact
positions.
4.4

Analogue Output Cable

Make the appropriate connections, ensuring that the correct polarity and the maximum load
specification is strictly observed.
4.5

Rear Connections:

4.6

Wiring the AMT

The AMT is a 2-Wire 4-20mA transmitter.
Typical electrical connection to generate a voltage output is shown below:
V o u t ma x = R s x 2 0 x 1 0 - 3
Note, the maximum value for Rs in this circuit can be calculated by the formula:
AM T
P i n 2
P i n 1
4 - 20mA Signal
Vout
SENSE
R s
RESISTOR
Page 3
+V s
P O W E R S U P P L Y
+ 1 2 V t o + 2 8 V
0V