Figure 22 – Remove the stop-pin spring from the slide

Reassemble the Condenser-Height Stop

Use a 1.5mm Allen wrench or driver to back out the
M3x4 stop-pin set screw so that the threads do not
protrude into the bore in the condenser slide (see
Figure
23).
Figure 23 – Back out the stop-pin set screw from the slide
Use a cotton swab and a suitable solvent (e.g., acetone)
to thoroughly clean the stop pin and the bore of the
condenser slide. Test the fit of the stop pin in the bore.
The pin should slide freely in the bore without any
noticeable friction or binding (see
Figure 24 – Test the fit of the stop pin in the bore
It is critical to the operation of the condenser-height
stop that this pin move freely in the bore in the
condenser slide. If there is noticeable friction or any
binding present, use a .1990" (i.e., 5.05mm) twist drill
Complete Teardown, Cleaning/Lubrication, and Reassembly of the Olympus BH2-CH Substage Assembly
Figure
24).
bit to clean the bore of any debris or internal burring
(see
Figure
25).
Figure 25 – Clean any debris or burring from the bore
Place the stop-pin spring into the bore in the condenser
slide (see
Figure
26).
Figure 26 – Reinstall the stop-pin spring into the bore
Place the M5x12 stop pin into the bore in the condenser
slide (see
Figure
27).
Figure 27 – Reinstall the stop pin into the bore
Use a 1.5mm Allen wrench or driver to tighten the
M3x4 set screw to secure the stop pin in position (see
Figure
28).
Revision 1