Problem
Dirt or dust is visible in the field
of view.
Visibility is poor
· Image is not poor.
· Contrast is poor.
· Details are indistinct.
· Image glares.
One side of image is blurred.
Image appears to waver.
Coarse adjustment knob is hard
to turn.
Image cannot be focused:
· Objective drifts down by itself.
· Fine adjustment knob slips.
28
Lamp
Cause
Common
Dirt/dust on the eyepieces.
Common
Dirt or dust on condenser top
lens.
Common
Dirt/dust on the specimen.
Common
A non-UIS/UIS2 objective is
used.
Common
Condenser height is too low.
Common
Aperture iris diaphragm is not
opened enough.
Common
Revolving nosepiece is not at-
tached properly.
Common
Correction collar on correction
collar-equipped objective is not
properly adjusted.
Common
Front lens of objective is dirty.
Common
Immersion oil is not being used
with an oil immersion objective.
Common
Immersion oil contains bubbles.
Common
Recommended immersion oil is
not used.
Common
Dirt/dust on specimen.
Common
Dirt/dust on condenser top lens.
Common
Objective is not correctly engaged
in light path.
Common
Stage is not correctly mounted.
Common
Revolving nosepiece is not at-
tached properly.
Common
Specimen is mounted upside
down.
Common
Revolving nosepiece is not
attached properly.
Common
Objective is not correctly engaged
in light path.
Common
Condenser is not properly
centered.
Common
Tension adjustment ring is
tightened excessively.
Common
You are trying to raise stage
without coarse adjustment knob
while pre-focusing lever is kept
locked.
Common
Tension adjustment ring is too
loose.
Remedy
Clean thoroughly.
Use only UIS2/UIS series objecti-
ves with this microscope.
Adjust correctly.
Open aperture iris diaphragm.
Push the slide dovetail all the way
until it is stopped.
While focusing, turn the correction
collar to find the best position.
Clean objective.
Use immersion oil.
Remove the bubbles.
Use an Olympus-designated
immersion oil.
Clean it.
Make sure that revolving nosepiece
clicks into place correctly.
Re-attach it.
Push the slide dovetail all the way
until it is stopped.
Mount specimen correctly.
Push the slide dovetail all the way
until it is stopped.
Make sure that revolving nosepie-
ce clicks into place correctly.
Center the condenser.
Loose ring.
Unlock pre-focusing lever.
Tighten ring.
Page
3
32
24
12
35
25
3
25
25
25
3
12
18
35
18
35
12
12
16
17
16