Olympus CK40 Manual de instrucciones - Página 13
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus CK40. Olympus CK40 39 páginas. Culture microscope
²
³
ƒ
Fig. 15
†
Fig. 16
Connecting the Lamp Cord Plug and
10
Power Cord
Do not subject the power cord to excessive force. Cables and cords
are more susceptible to damage when bent or twisted.
@
Make sure the main switch @ is set to " \ " (OFF) before connecting
the power cord. (Fig. 15)
1. Connect the lamp cord plug @ securely to the connector ² on the rear
|
of the microscope body. (Fig. 15)
Do not connect anything other than the lamp cord plug @ to the
connector ². Doing so will cause equipment failure.
Always use the power cord provided by Olympus. If no power cord
is provided, please select the proper power cord by referring to the
‡
section "PROPER SELECTION OF THE POWER SUPPLY CORD" at
the end of this instruction manual.
2. Be sure to set the input voltage selector ³ to the required voltage. Input
voltage is factory-set to 110 – 120 V or 230 – 240 V.
3. Insert the power cord plug | securely into the connector ƒ. (Fig. 15)
4. Insert the power cord plug † securely into the wall outlet ‡. (Fig. 16)
Connect the power cord correctly and ensure that the ground termi-
nal of the power supply and that of the wall outlet are properly con-
nected. If the equipment is not grounded, Olympus can no longer
warrant the electrical safety and performance of the equipment.
CK30/CK40
(Figs. 15 & 16)
8