Olympus CK40M FRAME Manual de reparación - Página 5

Navegue en línea o descargue pdf Manual de reparación para Microscopio Olympus CK40M FRAME. Olympus CK40M FRAME 38 páginas. Compact inverted metallurgical microscope

Olympus CK40M FRAME Manual de reparación
CK40M
• When CK40-MVR (mechanical stage) is mounted on the right side (seen from the
operation knob side) of the plane stage, CK40M-MS (stage mirror) cannot be mounted
on there.
• When it is combined with CK40M-CP (stage plate), the observation range depends on
the objective. In some cases, the CK40M-CP is pushed up by changing to a high
magnification objective. (High magnification)
Combination with MDPlan80X:
The narrow part in the clip hole (shape: tear drop) of the stage plate cannot be
observed. The part around φ15 mm in the clip hole of CK40M-CP is pushed up
by changing to the objective.
Combination with MDPlan50X:
When the clip hole (shape: tear drop) of the stage plate is positioned in the non-
breadthwise direction (when it is not combined with CK40M-MVR), the part
around φ15 mm in the clip hole of CK40M-CP is pushed up in some cases by
changing to the objective.
• When DP10 is combined with NFK2.5X, density may be uneven on one side of the field.
This is a restriction of DP10.
• When it is combined with a TV camera including a photo element, a spot flare may
occur on the TV monitor at a high-reflection mirror when the objective 50X or higher +
NFK2.5X or 3X are used.
(This is caused by surface reflection of the CCD surface and the NFK lens.)
• When it is combined with a TV camera including a 2/3 inch or larger photo element and
NFK1.67X, vigentting will occur at the four corners of the screen.
• When it is combined with CK40M-PO, the polarizer can be rotated more than 90° from
the crossed Nicol (total extinction) position to the index side, but less than 90° in the
opposite direction.
(3)
This model can be combined with the following objectives:
MDAch10X
MDAch20X
MDAch50X
MDPlan5X
MDPlan10X
MDPlan20X
MDPlan40X
MDPlan50X
MDPlan80X
This model cannot be combined with the following objectives: LB (long barrel) objectives
other than the above, SB (short barrel) objectives, and UIS objectives.
A. OUTLINE OF PRODUCT
A-2