Olympus CIX100 Manual de instrucciones - Página 26

Navegue en línea o descargue pdf Manual de instrucciones para Equipos de laboratorio Olympus CIX100. Olympus CIX100 46 páginas. Cleanliness inspector system
También para Olympus CIX100: Manual de configuración de sistemas (38 páginas), Manual de instalación (14 páginas)

Olympus CIX100 Manual de instrucciones
OLYMPUS CIX100 Cleanliness Inspector System
l
3. Select an observation method which is still undefined. If you start on top,
you will start with the new sample type without real color slider selected.
4. Make sure that the matching sample type is mounted. If not, carefully mount
the sample following the safety instructions in the chapter "Safety" on
page 9.
5. Click [Next>] to continue.
6. Skip the acquisition of the dark current correction image. Dark current
calibration does not need to be repeated here as it was already calibrated
during installation (i.e. keep the checkbox selected) and click [Next>].
7. Select all check boxes in the following dialog except [Shading Correction]
and [Exposure Correction] and click [Next >].
26
Each observation method is marked with a tick if the white balance
has been performed already and marked with a yellow exclamation
mark if the white balance has not yet been performed.