Olympus U-UCD8 Manual de instrucciones - Página 19
Navegue en línea o descargue pdf Manual de instrucciones para Equipos de laboratorio Olympus U-UCD8. Olympus U-UCD8 40 páginas. Universal condenser
8. Move the condenser up and down until uniform darkfield illumination is attained.
9. Open the field iris diaphragm to the extend that even brightness is attained.
# Keep eyes away from the eyepiece while changing the objective during darkfield observation or
changing from darkfield observation to another observation mode.
If the objective is changed or the turret is rotated to switch the darkfield observation to another observa-
tion mode, direct light may enter your eyes.
2-5 Simple Polarized Light Observation (KPO)
Applicable objective power
Power
Simple polarized
light (KPO)
* When the PlanApoN2X, PlanApo2X or UPlanSApo4X, UPlanApo4X objective, or Ph objective for phase
contrast observation is used in polarized light observation, the image contrast may be poorer than the
image observed with other objective magnifications.
}To perform simple polarized light observation, an analyzer (U-ANT or U-AN) is needed. If you have not
installed the fluorescence illuminator yet, install the illuminator by referring to the instruction manual of the
U-ANT transmitted light analyzer.
}When you have already installed the fluorescence illuminator in your system, attach the U-AN reflected
light analyzer by referring to the instruction manual of the reflected fluorescence attachment.
1. Rotate the turret to select the BF brightfield light path (with no optical element engaged).
2. Push in the polarizer handling knob to engage the polarizer into the light path.
3. Mount the objective to be used in the revolving nosepiece, and rotate the nosepiece to swing in the
objective.
2X
4X
10X
¦
¦
Top lens OUT
20X
40X
60X
¦
¦
¦
Top lens IN
U-UCD8
100X
¦
1 6