4. Rotate the polarizer rotation knob to achieve a perfectly black field and then tighten the clamping knob.
}The "crossed Nicol" (perfectly dark) position is located near the 0
5. Place the specimen on the stage and move the stage up and down to bring the specimen into focus.
6. Adjust the field iris diaphragm opening until it circumscribes the field of view.
7. Stopping down the aperture iris diaphragm may further increase the contrast of the image.
@
1 7
Fig. 4
o
General Precautions in Operation
1. If reflected light fluorescence microscopy in the U-excitation
mode is performed using a 10X or 20X objective with the
condenser engaged, flare may become prominent depend-
ing on the specimen condition.
If this occurs and if you do not need transmitted light DIC, either
lower the condenser or use the light shield sheet @ provided
with the reflected light fluorescence attachment. (Fig. 4)
2. When the polarizer rotation knob is set to the 0
dex, the polarizer and analyzer are nearly at the "crossed Nicol"
position. However, since a perfect "crossed Nicol" position can
hardly be achieved due to slight unavoidable positioning er-
ror of the intermediate tube, always perform fine adjustment.
3. When the top lens is swung out, the field of view become
obscured if the aperture iris diaphragm is stopped down (when
using a 2X or 4X objective).
4. Due to the large numerical aperture (N.A.), the U-DIC60, U-
DPO60S, U-DIC100 or U-DP100 DIC prism for 60X and 100X
objectives can also be used for brightfield observation.
position index.
o
position in-