9. Maintenance
Always disconnect the power cords for cleaning. When
cleaning the surfaces of the instrument and LED
transilluminator, use a damp cloth with mild detergent if
needed. Do not use any corrosive solutions that could
damage plastic.
When cleaning the gel tank, first separate the Power
Supply. Use water or neutral cleaner.
Use caution when cleaning the areas near the
platinum electrodes at the bottom edges of the gel
tank.
Do not submerge the power supply or LED transilluminator
in water.
Clean gel casting stands, gel tray and combs with water or
a neutral cleaner.
10. Troubleshooting
Problem
No display
No Migration
Abnormal
Electrophoresis
Distance
Abnormal heating of
buffer or gels
Key function failure
Test Equipment Depot - 800.517.8431 - 5 Commonwealth Ave, Woburn, MA 01801
Causes
Power not connected.
Switch failure. Fuse
blown.
Controller failure.
Power switch off, no
power, gel inserted
backwards
Incorrect input voltage
Buffer concentration
too high, voltage too
high
Control panel switch
failure
TestEquipmentDepot.com
14
Resolutions
Check power supply, Unplug and re-
attach power cord. Check fuse.
Contact Benchmark for repair.
Turn on power switch. Check that gel
is properly oriented (DNA will migrate
towards + electrode)
Contact Benchmark for repair.
Check main power supply.
Use 0.5X TAE/TBE buffer, 1X
QuickSilver Fast Running
reduce running voltage.
Contact Benchmark for repair.
Buffer,