Troubleshooting {continued)
Problem
Lid opened during run or
Error E1
magnet switch problem.
Lid not closed upon start or
Error E3
magnetic switch problem.
Electric current not flowing
Error E5
normally. Buffer
concentration too high.
Buffer concentration too high,
current too
Error E7
50W, or internal electronic
problem
Additional Electrophoresis Products from Accuris:
SmartGlow
E4S00-LD
E4S00-PS
QuickSilver
EB1000
EB1001
EB1002
Benchmark Agarose LE (Molecular Biology Grade)
A170S
1\1701
A1700
e
Please contact your Accuris distributor for pricing
r-r-1 l:IIC::
'-.._.._..._.,I
Jr7=5J""�LJrnE:rr-r=:;
Test Equipment Depot - 800.517.8431 - 5 Commonwealth Ave, Woburn, MA 01801
Causes
missing,
high,
power over
Reagents
ISmartGlow
Loading Dye with Safe Green Stain, 1.0ml
ISmartGlow
Safe Green Pre Stain, 1.0ml
Powdered Buffer Packs
OuickSilver TAE Buffer Powder, 100 pouches
QuickSilver TSE Buffer Powder, 50 pouches
OuickSilver Fast Running Buffer Powder, 100 pouches
Benchmark Agarose LE S00g
Benchmark Agarose LE lQQg
Benchmark Agarose LE, 2Sg
a division of
■--­
'-
TestEquipmentDepot.com
Resolutions
Seat lid properly.
magnet on lid ..
Seat lid properly.
magnet on lid.
Check proper buffer level in
the tank. Check buffer
concentration (1X or 0.5X
or
TAE or TSE is standard).
Check electrodes and
connections.
Check buffer concentration.
Replace electronic board.
15
Check
Check