Siemens 25 Manual de servicio - Página 6
Navegue en línea o descargue pdf Manual de servicio para Controlador Siemens 25. Siemens 25 12 páginas. Combination starters
También para Siemens 25: Piezas de repuesto y manual de instrucciones (6 páginas), Manual de solución de problemas (39 páginas)
8.6
Safety Checks
The user shall follow the following safety checks
8.6.1 Daily checks
Check the power cord. Do not use the unit if the power cord is damaged.
Check the exposure indicator and audible signal during exposure.
Check the locking mechanism for the support arm.
Check that the Single tank assembly and the arm system remain in the desired
position.
8.6.2 Monthly checks
Check the lifting and lowering movements of the support arm.
Check function of locking lever and cap screw on the X-ray tube.
Check the legibility of the labels according to those in section Location of Labels.
If any label requires to be replaced, please notify Siemens customer service.
8.7
Performance Check (Every Six months)
Check Counterbalancing of the support arm at each position.
Check the locking mechanism for the support arm.
Check Brake for proper functioning.
Take a sample exposure at 60 kV, 10 mAs.
Check for Radiation indication LED and the Audio indication in form of Buzzer.
NOTICE
Siemens Ltd. Med India
Annual Maintenance must be performed as per the
guidelines described in the Maintenance section.
Copyright © SIEMENS LTD. All rights reserved. For internal use only
Maintenance Instructions
Version 6.0
Service Manual
8-5