Siemens 25 Instrukcja serwisowa - Strona 6

Przeglądaj online lub pobierz pdf Instrukcja serwisowa dla Kontroler Siemens 25. Siemens 25 12 stron. Combination starters
Również dla Siemens 25: Części zamienne i instrukcja obsługi (6 strony), Instrukcja rozwiązywania problemów (39 strony)

Siemens 25 Instrukcja serwisowa
8.6

Safety Checks

The user shall follow the following safety checks

8.6.1 Daily checks

Check the power cord. Do not use the unit if the power cord is damaged.
Check the exposure indicator and audible signal during exposure.
Check the locking mechanism for the support arm.
Check that the Single tank assembly and the arm system remain in the desired
position.

8.6.2 Monthly checks

Check the lifting and lowering movements of the support arm.
Check function of locking lever and cap screw on the X-ray tube.
Check the legibility of the labels according to those in section Location of Labels.
If any label requires to be replaced, please notify Siemens customer service.
8.7

Performance Check (Every Six months)

Check Counterbalancing of the support arm at each position.
Check the locking mechanism for the support arm.
Check Brake for proper functioning.
Take a sample exposure at 60 kV, 10 mAs.
Check for Radiation indication LED and the Audio indication in form of Buzzer.
NOTICE
Siemens Ltd. Med India
Annual Maintenance must be performed as per the
guidelines described in the Maintenance section.
Copyright © SIEMENS LTD. All rights reserved. For internal use only
Maintenance Instructions
Version 6.0
Service Manual
8-5