Digitimer NeuroLog NL405 Width/Delay Manual del usuario - Página 4

Navegue en línea o descargue pdf Manual del usuario para Generador de impulsos Digitimer NeuroLog NL405 Width/Delay. Digitimer NeuroLog NL405 Width/Delay 5 páginas.

Users Manual
Rear Connection IN/OUT jumpers (on-board)
NL405 COMPONENT VIEW
JMP1
JMP2
The rear edge connector in the NL900 rack allows adjacent modules to connected together without the need of
external front panel leads.
This module has a choice of three output signals that can be connected to the rear connector for automatic
routing to the module on the immediate right. The Input signal can be disconnected from the output of the
module on the immediate left if it is inappropriate.
The position of these jumpers are changed by gently pulling them away from the PCB, then rotating them 90°,
180° or 270° before replacing them in two of the small gold board sockets. In the "Parked" or "N/C" position the
jumper is placed in only one gold socket and a parking hole.
INPUT - the upper jumper allows the Output signal from the module to the immediate left to be connected to the
input socket (IN), or it may be "parked" for "no-connection (N/C)".
OUTPUT - the lower jumper selects which of the three Output signals is passed to the Input of the module to
the immediate right, or it may be "parked" for "no-connection (N/C)".
Digitimer Ltd
NeuroLog™ System
Parked
Parked
Fig. NL405-3 : On-board IN/OUT jumpers
Page 4 of 5
Tel:+44 (0)1707 328347; Fax:...373153
IN
OUT
OUT
Website: www.digitimer.com
NL405
DELAYED
Copyright © 2003