Argos Technologies EVac Manual del usuario - Página 13

Navegue en línea o descargue pdf Manual del usuario para Equipos de laboratorio Argos Technologies EVac. Argos Technologies EVac 20 páginas.

Operation

Control knob allows smooth regulation of vacuum pressure. Suction power is indicated by
graduated marker on dial (see below).
Low Suction
Before starting suction, check that there is sufficient free volume in bottle for the amount
of liquid being aspirated.
CAUTION
Do not overfill bottle.
STOP
The EVac™ includes a safety filter which will stop fluid from reaching internal
mechanisms. If fluid or foam gets into filter, the filter must be changed immediately
or unit may no longer retain liquids. Note that hydrophobic filter holds back aqueous
solutions, whereas retention of apolar liquid is limited.
When working with dangerous substances, please maintain compliance with all
applicable safety instructions, national regulations (GLP, GMP, etc.), trade associations,
public health departments and factory inspectors.
When working with hazardous volatile substances or biohazardous agents, always place
the EVac™ within a laminar flow cabinet.
NOTE
i
If equipment does not operate correctly, please refer to the "Troubleshooting (FAQ)"
section on page 15.
Argos Technologies EVac™ — Operator's Manual
12
High Suction