Maintenance, Care and Safety Precautions
Maintenance
The EVac™ does not require any special servicing or maintenance. If a reduction in
suction power occurs, it may be caused by blockages in the tube and/or filter on the pump
side. This may be remedied by exchanging the affected tube and/or filter. When replacing
filter, remember that writing on filter should point away from device.
Care
The EVac™ may be cleaned with most common laboratory disinfectants, including
isopropanol.
NOTE
i
When using level detection, always ensure that the contact pins on bottle lid are kept
clean and dry.
DANGER
Always unplug the EVac™ from power supply before cleaning. Never use caustic
solutions (such as nitro thinners) to clean the EVac™.
Autoclaving
The EVac™ bottle, lids and tubes may be autoclaved to a maximum temperature of
121° C (249.8° F).
Refer to the autoclave manufacturer's guide for correct sterilization of liquids
and volumes.
To prevent damage, always disconnect quick connectors prior to autoclaving.
Loosely screw lid onto bottle during autoclave process to prevent heat from
STOP
deforming bottle lid.
The EVac™ Instrument, Mains Adapter, and Sensor Cable are not autoclavable.
To prevent generation and/or aspiration of corrosive vapors, use only non-caustic
STOP
chlorine-free agents for chemical sterilization.