Disabling the Radio-Link
Software to Save Battery
Power
Label
Typical Display
SEt-1**
SEt-2**
FiLt
SErch*
Addr*
bAud*
*
Change takes effect only after the unit has been switched off and then back on.
**
Used only in special software applications

Sleep Mode

10
The ED-2000plus software allows you to use the Radio-Link option. If the
Radio-Link option is installed but not being used, disable the radio portion of
the software to increase battery life. Also, if your system is not equipped with
the Radio-Link option, disable the software.
To disable the Radio-Link software see the SErch section in Table 5 below.
Notice that the change will not take effect until the ED-2000plus has been
switched off and then back on.
Description
100
Typically not used.
110
Typically not used.
2
Filter values (damping factors) on the load cell signal. Acceptable
values are from 0 (no damping) to 7 (maximum damping).
rS-485
Changes the serial communication link between RS-485 and radio
rAdio
link. Press F + TEST key to toggle between the two choices. Select
RS-232 communication by depressing F+ON/OFF at power up and
RS-485 and radio will be bypassed.
100
The communication address assigned to the device in multidrop
applications. Values should be between 33 and 255.
2400
Communication baud rate. Available values are 300, 1200, 2400,
4800, 9600, and 19,200 baud. System default is 2400 baud. Radio
communication at 2400 baud by default.
Table 5
Configuration Items
Sleep mode is a non-configurable battery saving feature of the ED-2000plus. It
functions all the time. If signal changes from the strain gages are not received
by the microprocessor, certain high drain functions are shut down. The
display does not change and any communication with the unit or signal from
the gages reactivates the unit instantly. The function of the sleep mode is
transparent to the user.