Nikon Eclipse E600 Manual de instrucciones - Página 18

Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Nikon Eclipse E600. Nikon Eclipse E600 21 páginas. Teaching head/multi teaching head
También para Nikon Eclipse E600: Manual de instrucciones (38 páginas), Manual de instrucciones (37 páginas), Manual de instrucciones (37 páginas)

Nikon Eclipse E600 Manual de instrucciones
IV

Troubleshooting

Focus differs among the main and secondary observer positions.
→ The diopter setting has not been adjusted correctly.
→ The components with different serial numbers are
put together.
The arrow is not visible when the pointer is used.
→ The arrow is outside of the view field.
→ The arrow color selector dial is in an intermediate
position.
→ The arrow brightness control is OFF.
→ The AC adapter is not connected.
→ The lamp is burned out or no lamp is mounted.
When using the pointer, the arrow and the specimen are not in simultaneous
focus.
→ The diopter setting has not been adjusted correctly.
The arrow is blurred when the pointer is used.
→ The arrow is too bright.
→ Adjust the diopter setting
correctly. (P.8)
→ Reassemble the components
with the identical serial
numbers to each other.
→ Use the arrow control stick and
bring the arrow into the view
field.
→ Turn the dial in the preferred
direction as far as it will go.
→ Turn the control clockwise.
→ Connect the AC adapter.
→ Mount a new lamp. (P.16)
→ Adjust the diopter setting
correctly. (P.8)
→ Turn the arrow brightness
control counterclockwise to
reduce the brightness.
17