Troubleshooting

Focus differs among the main and secondary observer positions.
→ The diopter setting has not been adjusted correctly.
→ The components with different serial numbers are
put together.
The arrow is not visible when the pointer is used.
→ The arrow is outside of the view field.
→ The arrow color selector dial is in an intermediate
position.
→ The arrow brightness control is OFF.
→ The AC adapter is not connected.
→ The lamp is burned out or no lamp is mounted.
When using the pointer, the arrow and the specimen are not in simultaneous
focus.
→ The diopter setting has not been adjusted correctly.
The arrow is blurred when the pointer is used.
→ The arrow is too bright.
→ Adjust the diopter setting
correctly. (P.8)
→ Reassemble the components
with the identical serial
numbers to each other.
→ Use the arrow control stick and
bring the arrow into the view
field.
→ Turn the dial in the preferred
direction as far as it will go.
→ Turn the control clockwise.
→ Connect the AC adapter.
→ Mount a new lamp. (P.16)
→ Adjust the diopter setting
correctly. (P.8)
→ Turn the arrow brightness
control counterclockwise to
reduce the brightness.